| SINCLAIR COMMUNITY
COLLEGE |
Section: T 803 om-88 I |
| QET 134 COURSE -QUALITY
LAB |
Issued by: Mary E. Haggard |
| Subject: PUNCTURE TEST
OF CONTAINERBOARD |
Approved: James Houdeshell
Instructor |
1.0 PURPOSE
- This standard
provides a guideline for puncture test of containerboard.
- The significance
is to simulate damage that would (or could) occur when product
is in use environment.
2.0 SCOPE
- This Standard
applies to Sinclair Community College Quality Lab.
- This test measures
the energy required to puncture containerboard or corrugated
board with a pyramidal point affixed to a pendulum arm.
- The energy to
tear the material and the energy to bend it out of the way of
the point are the two major components that make up this test.
3.0 RESPONSIBILITIES
- The Quality (Department)
Manager has the responsibility to ensure compliance with this
specification.
- The Quality (Department)
Manager has the responsibility to ensure proper training of equipment
and safety regulations.
- The Quality department
is responsible for performing the tests and recording the results.
4.0 REFERENCES
Beach, R. L., "
Puncture Testing of Box Board," Paper Trade J. 108
(5): 30 (1939).
Wells, S.D., Kaplan,
A.B., and Ayers, L.R., " Tests on Fiber Containers : The
Beach Puncture Tester,"Paper Trade J. 116(13):
29 (1943).
Van Den Akker,
J. A., and Wink, W. A., " Instrumentation Studies: LX. The
Beach Puncture Tester," TAPPI 33(12): 114A
(1950).
5.0 EQUIPMENT
Puncture Test Apparatus
consisting of:
- Pendulum, with
an arm of a 90 degree angle.
- Puncture point,
affixed to end of arm with a shape of a right triangle pyramid
25 +/-0.7mm high, with leading edges rounded, base edges left
sharp, and the point to 1.57 +/- 0.05mm(0.062+/-0.002 in.) radius.
- Puncture Point
Collar, loose in fit to avoid friction around base.
- Trigger, or latch
to hold pendulum in n horizontal plane.
- Clamping jaws,
spring actuated to hold specimen in a horizontal plane.
- Pointer and scale,
to indicate the arc through which the pendulum travels after
passing through the specimen. Scale graduations correspond to
weights used.
- Arguementing weights,
giving scale values up to least 38.87 J (1300 units of energy).
- Loose plate, inserted
between the clamping jaws to define the test area securely attached.
*Note: this plate is removed before checking the zero reading
of the apparatus.
CALIBRATION
& ADJUSTMENT
If apparatus is
not under a calibration system then follow the next steps to calibrate
apparatus:
- Attached apparatus
to concrete base or any other foundation of convenient height,
and bolt down.
- Pendulum Friction- Remove all weights and
weight support stud, put puncture collar in place on puncture
head. Rotate pointer to below zero mark latch and release pendulum.
Pendulum should swing a minimum of 8 minutes before coming to
rest. If it does not clean and lubricate the bearing
- Zero reading. - Remove all weights
and weight support stud, put puncture collar in place on puncture
head. Allow the pendulum to hang vertically on its pivot shaft.
With adjustment screw align the point exactly on full scale.
Set point approximately 25mm (1in.) above zero on the scale raise
and latch pendulum. Release pendulum. The scale reading should
read zero. If it does not, then loosen the four setscrews that
position the latch tongue and repeat procedure until zero reading
is achieved.
- Pointer Friction. - Latch the bare pendulum
with only the collar attached to the head. Remove clamping plate,
rotate the pointer to full-scale reading, release pendulum allowing
to swing and push the pointer further. Adjust the pointer friction
by the screws in the stud projecting from the pointer hub, so
that additional movement is not more than two units. Recheck
pointer at mid-scale. If friction is reduced too much the pointer
will move or "bounce" below the position indicated
in the second step.
- Puncture point
collar adjustment- Place
collar in same position each time on the puncture head. Mark
collar if necessary to achieve proper location each time. An
adjustable pin determines the force needed to dislodge the collar.
The force is 1.0+/- 0.15N (3.6+/-0.5oz) pull when the collar
is in a horizontal position. When collar is attached the puncture
head extends at least 0.38mm(0.015in.) beyond the outer edge
of the collar. Keep flange angle at approximately 135 degrees.
- Clamp load-
A force
of 245+/- 22N (55+/- 5 lb) applied downward from the middle of
the lower plate should open the clamp 6mm (1/4in). Tighten or
loosen the wing nuts at bottom of spring tube to adjust the pressure
if necessary. Using test specimen of the thinnest strips, check
uniformity of clamping, by placing the strips between the jaws
at various points. If any of these can be easily removed adjust
the jaw alignment until clamp load is uniform.
- Center of gravity-
The alignment
of the center of gravity may be most conveniently checked by
removing the weights and studs from the pendulum, allowing pendulum
to come to its rest position (with collar attached at the puncture
point), and gently rotating the pointer clockwise until it just
touches its adjusting screw. NOTE the pointer position,
attach stud and weights to the pendulum, and repeat above operations.
If the center of gravity lies in the proper plane, no shift of
the pointer will occur. If pointer shifts, the center of gravity
is incorrect. Correct it by shifting the balancing weights in
or out on its screw. The weight are adjusted and stacked in place
at manufacturer, they should only be moved when it is 100% certain
that the center of gravity is off. If adjustment is appeared
to be necessary then one should check first the pendulum with
the combination of weights and by rotating weights to make sure
they are concentric before returning the pendulum to the manufacturer
for adjustment.
- Loose plate-Secure the loosed plate
in a vertical projection position beneath the upper clamping
jaws such that the bottom edge of upper jaw is 31.8mm
(1 and 1/4 in.) from the base of puncture head.
- Puncture point-
Be sure
that the three corners at the base of the puncture point are
equidistant (within 1.5mm (1/16 in.) below the upper clamping
jaw when the base of the puncture point is 31.8mm(1 and 1/4 in.)
below the upper clamping jaw; that its apex does not deviate
in plane more than 1.5mm (1/16in.) from pendulum. And that the
edge farthest from the axis of rotation is parallel to that axis.
Keep puncture protected from abuse, free from nicks and scratches.
Clean pointer after each use with a degreasing agent.
SAMPLING AND
TEST SPECIMENS
- Sample the board
in accordance with TAPPI T 400 " Sampling and Accepting
a Single Lot of Paper, Paperboard, Fiberboard, of Related Product.
- Select 3 representative
specimens at least 300 x 300 mm (12 x 12 in.)
- All test are to
be made at least 95mm (3 and 3/4 in.) from an edge, score line,
or imperfection.
CONDITIONING
- Precondition and
condition the specimens and test in atmosphere specified in TAPPI
T 402 " Standard Conditioning and Testing Atmospheres for
Paper, Board, Pulp hand sheets, and Related Products."
PROCEDURE
Place puncture point
collar on base of the point and set the pointer about 25mm (1in.)
above expected reading. Raise and latch pendulum.
9.2. Scales plate
for test is attached to apparatus and the weights to be used for
each scale are indicated on the scale plate
9.3 For single,
double, & triple wall corrugated, and solid fiberboard
4 punctures comprise
a set. One set constitutes one test
- Orientation of
a set :
- Orientation 1.
Parallel, with one surface down.
- Orientation 2.
Parallel with other surface down.
- Orientation 3.
Perpendicular, with one surface down
- Orientation 4.
Perpendicular, with other surface down.
9.4 For uncombined
sheets or single- faced corrugated boards
- 2 punctures comprise
a set. One set constitutes one test
- Orientation of
a set:
- Orientation 1.
Parallel with smooth surface down.
- Orientation 2.
Perpendicular with smooth surface down.
9.5 Clamp each specimen,
the release the pendulum by pushing the latch handle to the left.
When pendulum completes its swing note the reading to the nearest
one-half division on the proper scale. * Note it is not advisable
to perform a test giving a reading within the upper or lower quarter
of the scale. Use another scale in such cases.
9.6 For lightweight
material use the same procedure as above except use the bare pendulum
without additional weights and stud, releasing the pendulum from
a 45degree position and the auxiliary low scale.
REPORT
- Report the average
of two test sets in scale units of three significant figures.
- Avg. of Test 1
+ Avg. of Test #2 Divided by 2 = Total Test Average.
PRECISION
- The repeatability
as defined in TAPPI T 1206 " Precision Statement for Test
Method " is 5.25%.
- The values of
repeatability have been calculated as stated in step 9.3-9.4
for a test result which is the average of a set (four individual
test)
- The reproducibility
of this method is not known.
|
REV.
LTR. |
REV.
DATE |
DESCRIPTION
OF CHANGE |
|
Original |
11/11/01 |
Original Issue |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|